User ManUalRaySafe Solo
R/F MEASUREMENTUsing R/F, RAD, Dose or DENT version.For best accuracy, center the selected sensor field (R/F low, R/F high, RAD, DENT low or DENT high)
2. MEASURE MODEAfter an oset adjustment, the instrument is in MEASURE MODE. The displayed values are updated after each exposure, or continuously aft
All values are stored in a non-volatile memory and are valid until manually changed. At start up (after the battery status information), valid Trig de
MAMMOGRAPHY MEASUREMENT Place the detector along the X-ray tube axis (with the cable end pointing towards the X-ray machine). Make sure that the cente
Dose rate Gy or R per s, m or hTime ms, sHVL mm Al (only with optional HVL kit)3. SETUP MENUPress SELECT to enter the SETUP MENU (from MEASURE MODE) a
Auto Power ONever, or after 5, 20 or 60 minutes of inactivity.Info Displays the serial number (S/N) and firmware version of the connected detector and
CT MEASUREMENTThe RaySafe Solo CT detector is an ion chamber with a 10 cm active length and a built-in bias voltage. The center and the edges (+ 5 cm
1. SENSOR MENUCT Dose Ion chamber sensorPress STEP to step between sensors and SELECT to select a sensor.NOTE! When turning on the base unit with th
Calc. delay Defined as the dead time after end trig, before data is calculated. Default is 0.5 s but it is recommended to use a longer delay when measu
LIGHT MEASUREMENTThe RaySafe Solo light detector is designed to measure luminance on LCD and CRT monitors, as well as light boxes, and illuminance for
© 2013.01 Unfors RaySafe 5001053-CAll rights are reserved. Reproduction or transmission in whole or in part, in any form or by any means, electronic,
Press STEP when in MEASURE MODE to freeze and move the measured value to the second line of the display. A new press on STEP will move a new measured
All values are stored in a non-volatile memory and are valid until manually changed.(Factory settings in bold.)Oset adjustment To ensure the highest
mA/mAs MEASUREMENTNOTE! It is possible to measure mAs simultaneously with other measurements with RaySafe Solo R/F. The instrument will trig on rad
2. MEASURE MODEAfter an oset adjustment, the instrument is in MEASURE MODE. The displayed values are updated after each exposure or continuously afte
Info Displays the serial number (S/N) and firmware version of the detector and the RaySafe Solo base unit and the calibration date for the selected sen
PASSIVE mAsIt is possible to measure mAs simultaneously when using the RaySafe Solo R/F with mAs. The instrument will trig on radiation, not on the cu
MEASUREMENT DEFINITIONSFigure 1 illustrates the RaySafe Solo parameter definitions available on a typical waveform. In most situations the default sett
The dose is accumulated from the start trig point plus Trig delay until the end trig.Figure 2. Trig delay activatedNOTE! The trig level setting (25,
DOSE RATEThe RaySafe Solo calculates dose rate as dose/exposure time if the exposure time is less than 6 seconds.After an exposure longer than 6 secon
Figure 4. kV and kVpTable: Active compensation on kV measurementsSENSOR FILTRATION kV/KVpR/F low Max. 15 mm Al kVR/F high RAD Max. 15 mm Al kVp (when
CONTENTSINTRODUCTION ... 5Versions ...
Figure 5. 75 % trig level activated (figure also applies to 25 and 50 % trig level settings). Please note that the trig level setting aects the do
HVL is an indication of the beam quality and is defined as the amount of Al filtration, measured in mm, needed to reduce the dose in half. HVL is kVp de
RAYSAFE SOLO PC KITThe optional RaySafe Solo PC kit contains:• RaySafe Xi View software• serial cable• serial to usb converter• optional: bluetoot
TRANSFER DATA TO A COMPUTERTo view the waveform in the RaySafe Xi View and/or transfer measured data to a PC, connect a serial cable or Bluetooth modu
7. Start RaySafe Xi View and go to File – Settings and enter the correct COM port. Press the Start button and “Started waiting for data” will appear
ACCESSORIESTHE RAYSAFE FLEXI STANDThe RaySafe flexi stand can be used in many dierent ways. Assemble it to fit your needs.THE RAYSAFE OPG HOLDER1. If
4. Expose and align the fluorescent line with the preferred black line (selected sensor) on the alignment lid. Soften the lighting if needed and compl
TIPS AND TRICKSARRANGEMENT OF MEASUREMENTMake sure nothing is in the X-ray field when measuring with the RaySafe Solo.The fact that the dose is proport
VERTICAL DETECTOR PLACEMENTWhen measuring on a mammography machine with a horizontal beam, such as MultiCare Platinum, let the pressure from the compr
DISPLAY MEANINGt<delay The exposure time is not long enough to calculate a value.Ofl. Overflow, i.e. calculated value is numerically higher than can
Setup of Bluetooth communication ... 33Getting help ...
I sometimes get readings which are far o expected when measuring with RaySafe Solo DENT on panoramic X-ray machines, why?Most likely the detector’s s
What happens if the mA/mAs cable is connected with the wrong polarity?The RaySafe Solo will not show any mA/mAs readings. As long as the mA/mAs cable
The RaySafe Solo detector end trigs during exposure.Make sure to select an appropriate Calc. delay if using the instrument with pulsed generators, or
WARRANTY, SERVICE AND SUPPORTUnfors Raysafe warrants to the original product purchaser that each product it manufactures will be free from defects in
DISPOSALThe Directive on Waste Electrical and Electronic Equipment (WEEE)This symbol, found on the product or on its box, indicates that this product
INTRODUCTIONVERSIONSThis user manual covers the following articles:base unit version: 8221011 detector version: 82220158221022 82220168222017822201882
THE RAYSAFE SOLOThe RaySafe Solo consists of a base unit and a detector.Communication between detector and base unit is purely digital, thereby minimi
EASY TO USEThe RaySafe Solo is very easy to use.The built-in active compensation automatically applies corrections for dierent beam qualities, filtrat
GETTING STARTED1. Connect the detector to the base unit with one of the two (2 and 10m) cables. 2. Position as required for the detector, see the fo
BATTERY AND CHARGERThe RaySafe Solo base unit has a chargeable 7.4 V Li-ion battery, providing approximately 20 – 40 hours of operation (depending on