SETTING: STOP DELAY
The stop delay setting defines for how long the instrument will wait for more radiation to include in the same
measurement.
Use a longer stop delay when measuring on a pulsed fluoroscopy system or to get a prepulse recorded into
the same measurement as the following ordinary exposure.
Stop delay setting longer than the time between pulses (t>T), will result in one long measurement. Note that
you have to wait as long as the stop delay setting (t) after the last exposure before the measurement ends
and values will be displayed.
Stop delay setting shorter than the time between pulses (t<T), will result in many short measurements (one
for each pulse).